URI | http://purl.tuc.gr/dl/dias/7A6E8713-7C8F-4A61-83E5-1A1F992A734B | - |
Αναγνωριστικό | https://doi.org/10.1016/j.aca.2007.03.066 | - |
Γλώσσα | en | - |
Μέγεθος | 8 pages | en |
Τίτλος | Microwave assisted headspace single-drop microextration of chlorobenzenes from water samples | en |
Δημιουργός | Elefteria Psillakis | en |
Δημιουργός | Claudia E. Domini | en |
Δημιουργός | Lorena Vidal | en |
Δημιουργός | Antonio Canals | en |
Δημιουργός | Nuria Gran | en |
Εκδότης | Elsevier | en |
Περίληψη | Aone-step and in-situ sample preparation method used for quantifying chlorobenzene compounds inwater samples has been developed, coupling
microwave and headspace single-drop microextraction (MW-HS-SDME). The chlorobenzenes in water samples were extracted directly onto an
ionic liquid single-drop in headspace mode under the aid of microwave radiation. For optimization, a Plackett–Burman screening design was
initially used, followed by a mixed-level factorial design. The factors considered were: drop volume, aqueous sample volume, stirring speed, ionic
strength, extraction time, ionic liquid type, microwave power and length of the Y-shaped glass-tube. The optimum experimental conditions found
from this statistical evaluation were: a 5 L microdrop of 1-hexyl-3-methylimidazolium hexafluorophosphate exposed for 20 min to the headspace
of a 30mL aqueous sample, irradiated by microwaves at 200W and placed in a 50mL spherical flask connected to a 25 cm Y-shaped glass-tube.
Under the optimised experimental conditions, the response of a high performance liquid chromatographic system was found to be linear over
the range studied and with correlation coefficients ranging between 0.9995 and 0.9999. The method showed a good level of repeatability, with
relative standard deviations varying between 2.3 and 8.3% (n = 5). Detection limits were found in the low gL−1 range varying between 0.016
and 0.039gL−1. Overall, the performance of the proposed method demonstrated the favourable effect of microwave sample irradiation upon
HS-SDME. Finally, recovery studies from different types of environmental water samples revealed that matrix had little effect upon extraction | en |
Τύπος | Peer-Reviewed Journal Publication | en |
Τύπος | Δημοσίευση σε Περιοδικό με Κριτές | el |
Άδεια Χρήσης | http://creativecommons.org/licenses/by/4.0/ | en |
Ημερομηνία | 2015-09-24 | - |
Ημερομηνία Δημοσίευσης | 2007 | - |
Βιβλιογραφική Αναφορά | L.Vidal , C. E. Domini , N. Grane , E. Psillakis , A. Canals , "Microwave assisted headspace single-drop microextration of chlorobenzenes from water samples " ,Anal. Chim. Acta ,vol. 592 ,no. 1 ,pp.9–15.2007. doi:10.1016/j.aca.2007.03.066 | en |